# Station Guide — Peak Intensity
“Peak Intensity” is the maximum pixel value measured in the target image.
Practical use: It provides a simple indication of major changes in illumination or target response.
A decreasing value can indicate:
• insufficient illumination;
• partial obstruction;
• contamination or reduced reflector response; or
• a missed or weakened illumination flash.
Very high values, particularly when they repeatedly reach the maximum available image
intensity, can indicate clipping or saturation.
Because Peak Intensity depends on a single pixel, it is sensitive to local variations and should
preferably be interpreted together with Flux and SNR.
Source : AIMsight Station Guide, rév. 1.5