Station Guide — Peak Intensity

“Peak Intensity” is the maximum pixel value measured in the target image. Practical use: It provides a simple indication of major changes in illumination or target response. A decreasing value can indicate: • insufficient illumination; • partial obstruction; • contamination or reduced reflector response; or • a missed or weakened illumination flash. Very high values, particularly when they repeatedly reach the maximum available image intensity, can indicate clipping or saturation. Because Peak Intensity depends on a single pixel, it is sensitive to local variations and should preferably be interpreted together with Flux and SNR.

Source : AIMsight Station Guide, rév. 1.5