# Station Guide — Exposure and Image Preparation
|
||||
|
||||
• Exposure: Aim for high contrast in the marker or reflector regions. Ideally, the reflector
|
||||
images should be bright, with brightness values above about 200 when hovering over them,
|
||||
and surrounded by a dark background. This is usually achieved with short exposure times,
|
||||
typically in the 0.1 ms to 10 ms range. When in doubt, use a shorter exposure time.
|
||||
• Plot brightness slider: adjusts only the display brightness of the image plot to help you
|
||||
assess exposure. It does not change the actual camera exposure time.
|
||||
• Recapture: clicking “Recapture” triggers a new image acquisition using the latest settings
|
||||
and updates the displayed image-processing zones and their names.
|
||||
• Image Type: for dynamic acquisitions using reflective targets as position markers, select
|
||||
“Single with infrared light”. The other options are mainly intended to optimize image
|
||||
acquisition in quasi-static monitoring scenarios. The differential image shows a contrastenhanced view of the reflectors in the scene and is used by the displacement-measurement
|
||||
algorithms by default.
|
||||
“Single with infrared light” captures one infrared-illuminated image per measurement and is
|
||||
best suited to dynamic measurements. “Differential” combines several images to increase target
|
||||
contrast and improve quasi-static tracking. In difficult lighting conditions or with targets that are
|
||||
not sufficiently reflective, use Differential mode with a higher Stacking value, typically between
|
||||
20 and 40. Higher stacking improves detection range but reduces the effective measurement
|
||||
rate.
|
||||
|
||||
Source : AIMsight Station Guide, rév. 1.5
|
||||