# Station Guide — Image Mode and Target Preset
• Single with infrared light: captures one infrared-illuminated image per measurement and is
generally suited to dynamic measurements requiring a higher acquisition rate.
• Differential: combines multiple images to improve the contrast of reflective targets and is
generally suited to quasi-static monitoring or difficult detection conditions.
Pattern Preset selects the target configuration used to calculate displacement. When (use active
preset) is selected, the schedule uses the preset active at execution time. A later change to the
active preset may therefore change which targets the schedule processes. Select a named
preset when the schedule must always use a fixed target configuration.
CAUTION
Before enabling an automatic schedule, verify that the reference patterns have been saved
correctly and that the reference image represents the required displacement zero.
Source : AIMsight Station Guide, rév. 1.5